Total SO2 content determination in shrimp paste by modified Monnier-Williams Method

MultiDist and BasicDist: Determination of total sulfur dioxide content in shrimp-paste by modified Monier-Williams method

Total SO2 content determination in shrimp paste by modified Monnier-Williams Method

Total SO2 is defined as “the total of all the various forms of sulfur dioxide present in the sample, either in the free state or combined with their constituents”. Its determination in solid samples poses a challenge because SO2 is withheld in adducts and is released slowly during boiling. Recognized results are obtained by means of the generally accepted Optimized Monier-Williams method according to AOAC 990.28 which suggests the use of nitrogen-assisted boiling of the acidified sample to release the Total SO2. The method presented in this study provides equally quantitative results with much less time required per sample and is based on steam distillation followed by pH titration.

In this application note recovery results for a widely used reference standard HMS, and a certified sample of shrimp paste (41±9.7ppm) for total sulfite analysis are presented. A representative concentration of 1 mg SO2 /sample tube is selected to demonstrate the robust performance of modified Monier-Williams method by steam distillation. 

Registration für Download

Land
Wir möchten Ihnen in Zukunft weitere relevante Informationen zusenden, z. B. Anwendungshinweise, Leitfäden, Webinar- und Seminareinladungen, die für Sie von Interesse sein könnten. Daher möchten wir Sie um Ihr Einverständnis bitten, mit Ihnen in Kontakt zu bleiben. Über Ihre Rechte und wie wir Ihre personenbezogenen Daten nutzen und verarbeiten, können Sie sich in unserer Datenschutzerklärung informieren. 

Ähnliche Applikationen